Reference Models for Receptor and other Tracers

Principle of Reference Models

The measurement and analysis of the blood samples for kinetic modeling is an invasive and demanding procedure. Therefore, methods have been developed to obviate the need for invasive blood sampling. The solutions found replace the arterial input curve by an indirect input curve, namely the time activity curve of some reference tissue. Therefore they are called reference methods. Reference methods are not able to provide a full kinetic analysis. However, assuming certain relations between the kinetics of the tissue of interest and the reference tissue, they can provide valuable measures of interest.

Model Structure

Most of these reference methods are dedicated to reversibly binding neuroreceptor tracers. A reference tissue must be found which is devoid of receptors, and then it is assumed that the distribution volume of the non-displaceable compartment (free tracer in tissue and non-specific binding) is the same among the tissues. Under these assumptions a measure of the receptor concentration called binding potential (BP) can be calculated from the two time-activity curves.

The reference methods differ in their mathematical approaches, and they show substantial differences with regard to noise sensitivity and processing speed. They are described in the following sections. The compartment models are based upon the following configuration:

In the model equations C'(t) represents the TAC from the reference region (k3=0 in the 2-tissue compartment model), and C(t) the TAC from a receptor-rich region (k3>0).

PXMOD Implementations

The measurements required for the receptor reference methods are a dynamic PET or SPECT acquisition without the need for external blood sampling. The operator must delineate a reference region devoid of receptors (e.g. frontal cortex for D2 receptors). For the model preprocessing step he also must delineate a receptor-rich region (e.g. basal ganglia for D2 receptors). The model then applies the reference model to the TACs from both regions and presents the results to the user for inspection. If the result is not satisfactory, the user may change some of the parameters and try the fit again. At the end of preprocessing some parameters such as the regression start time t* or k2' have been determined which will be used for the pixel-wise fits. The pixel-wise calculations result in BP maps. To avoid meaningless values which mix up the display it is recommended to restrict the accepted BP values to a reasonable range.